YES*FOR THIS SAMPLE

The result is generally acceptable.

Total Gage R&R is 1.9% of study variation and ndc is 74. Measurement variation is small relative to the selected part range. The asterisk matters: this five-part demonstration does not, by itself, approve the system for production use.

TOTAL GAGE R&R1.9%Generally acceptable band
REPEATABILITY1.7%Largest gage component
REPRODUCIBILITY0.9%Operator component
NDC74Strong sample discrimination

What supports “acceptable”

  • Total GRR is below the common 10% guideline.
  • Part-to-part variation clearly exceeds gage variation.
  • ndc is well above the usual minimum of 5.
  • The two operator means differ by only 0.0032 mm.

What still needs verification

  • The five parts represent the actual process range.
  • Measurement order was randomized.
  • Operators used the normal production method.
  • Resolution, calibration, fixture, and environment are adequate.
  • %Tolerance is acceptable when specification-based decisions matter.

The study data

Every operator measures every part twice. Values are in millimetres.

PartOperator A · Trial 1Operator A · Trial 2Operator B · Trial 1Operator B · Trial 2Part mean
110.01810.02110.02410.02010.02075
210.14610.15010.15310.14910.14950
310.27810.27410.28110.27610.27725
410.40710.41210.41510.41010.41100
510.53610.53210.54110.53710.53650

Why the result looks strong

The observed variation is overwhelmingly part-to-part. Repeatability is the larger measurement component, but it is still small beside the selected part range.

Total Gage R&R
1.9%
Repeatability
1.7%
Reproducibility
0.9%
Part-to-part
100%

How to read the familiar %GRR guideline

Below 10%Generally considered acceptable.
10% to 30%May be acceptable depending on risk, cost, and application.
Above 30%Measurement system usually requires improvement.

Do not game the part range. A wider part range can reduce %Study Variation and inflate ndc. Select parts that represent the real process variation; when conformance to specifications matters, also review %Tolerance.

Calculation trail

ComponentResultInterpretation
Equipment Variation (EV)0.019152 mmRepeatability under the study conditions
Appraiser Variation (AV)0.009987 mmOperator-to-operator component
Total Gage R&R0.021600 mmCombined measurement variation
Part Variation (PV)1.139808 mmVariation between selected parts
Total Variation (TV)1.140012 mmCombined study variation

For the complete Average & Range calculation sequence and constants, see the detailed calculation article.

Try the same study yourself

The free browser preview is preloaded with this dataset. Calculate it, change individual readings, and observe how repeatability, reproducibility, Total GRR, and ndc respond.

Bottom line

For this sample, yes: the measurement system appears capable of distinguishing the selected parts. For an actual approval decision, confirm the sampling, study execution, application risk, tolerance contribution, and required customer or company method. A good number is evidence—not a substitute for a valid study.